Agilent Technologies System DC N5700 User Manual

Agilent Technologies  
System DC Power Supply  
Series N5700  
User’s Guide  
A
Do Not Operate in an Explosive  
Atmosphere  
Do not operate the instrument in the  
presence of flammable gases or fumes.  
Safety Symbols  
Safety Notices  
Direct current  
The following general safety precautions  
must be observed during all phases of  
operation of this instrument. Failure to  
comply with these precautions or with  
specific warnings or instructions  
elsewhere in this manual violates safety  
standards of design, manufacture, and  
intended use of the instrument. Agilent  
Technologies assumes no liability for the  
customer's failure to comply with these  
requirements.  
Alternating current  
Do Not Remove the Instrument  
Cover  
Only qualified, service-trained personnel  
who are aware of the hazards involved  
should remove instrument covers. Always  
disconnect the power cable and any  
external circuits before removing the  
instrument cover.  
Both direct and alternating  
current  
Three phase alternating  
current  
Earth (ground) terminal  
Protective earth ground  
terminal.  
General  
Do Not Modify the Instrument  
Frame or chassis terminal  
Do not use this product in any manner  
not specified by the manufacturer. The  
protective features of this product may be  
impaired if it is used in a manner not  
specified in the operation instructions.  
Do not install substitute parts or perform  
any unauthorized modification to the  
product. Return the product to an Agilent  
Sales and Service Office for service and  
repair to ensure that safety features are  
maintained.  
Terminal is at earth  
potential.  
Neutral conductor on  
permanently installed  
equipment  
Before Applying Power  
Verify that all safety precautions are  
taken. Make all connections to the unit  
before applying power. Note the  
instrument's external markings described  
under "Safety Symbols"  
In Case of Damage  
Line conductor on  
permanently installed  
equipment.  
Instruments that appear damaged or  
defective should be made inoperative and  
secured against unintended operation  
until they can be repaired by qualified  
service personnel.  
On supply  
Off supply  
Ground the Instrument  
This product is a Safety Class 1  
CAUTION  
instrument (provided with a protective  
earth terminal). To minimize shock  
hazard, the instrument chassis and cover  
must be connected to an electrical  
ground. The instrument must be  
connected to the ac power mains through  
a grounded power cable, with the ground  
wire firmly connected to an electrical  
ground (safety ground) at the power  
outlet. Any interruption of the protective  
(grounding) conductor or disconnection of  
the protective earth terminal will cause a  
potential shock hazard that could result in  
personal injury.  
Standby supply. Unit is not  
completely disconnected from  
ac mains when switch is off  
A CAUTION notice denotes a hazard.  
It calls attention to an operating  
procedure, practice, or the like that, if  
not correctly performed or adhered to,  
could result in damage to the product  
or loss of important data. Do not  
proceed beyond a CAUTION notice  
until the indicated conditions are fully  
understood and met.  
In position of a bi-stable push  
switch  
Out position of a bi-stable  
push switch  
Caution, risk of electric shock  
Caution, hot surface  
WARNING  
A WARNING notice denotes a  
hazard. It calls attention to an  
operating procedure, practice, or the  
like that, if not correctly performed  
or adhered to, could result in  
personal injury or death. Do not  
proceed beyond a WARNING notice  
until the indicated conditions are  
fully understood and met.  
Fuses  
For continued protection against fire,  
replace the line fuse only with a fuse of  
the specified type and rating (normal  
blow, time delay, etc.). Do not use  
repaired fuses or short-circuited  
fuseholders. To do so could cause a  
shock or fire hazard.  
Caution, refer to  
accompanying  
documents  
Series N5700 User’s Guide  
3
In This Book  
This User’s Manual contains the operating instructions, installation  
instructions, and specifications of the Agilent Technologies Series  
N5700 750W and 1500W System DC Power Supplies. Specific  
chapters in this manual contain the following information:  
Quick Reference – Chapter 1 is a quick reference section that  
helps you quickly become familiar with your Agilent N5700  
power supply.  
Installation – Chapter 2 describes how to install your power  
supply. It describes how to connect various loads to the output. It  
discusses remote sensing as well as parallel and series operation.  
Operating the Power Supply Locally – Chapter 3 describes how to  
operate the power supply from the front panel and from the  
analog connector on the rear panel. It also includes a turn-on  
check-out procedure to verify the unit is operating properly.  
Operating the Power Supply Remotely – Chapter 4 describes how  
to configure the remote interfaces. It also gives a brief overview  
of the SCPI command structure and basic programming concepts.  
Language Reference – Chapter 5 describes all of the SCPI  
programming commands.  
Programming Examples – Chapter 6 provides Visual BASIC  
example programs that illustrate some common applications.  
Specifications – Appendix A describes specifications and  
supplemental characteristics.  
Verification and Calibration Procedures – Appendix B explains  
the verification and calibration procedures.  
Service – Appendix C describes what to do if your unit requires  
service.  
Compatibility – Appendix D documents the compatibility  
commands of the Agilent 603xA power supplies that are  
supported by the Agilent N5700 power supplies.  
You can contact Agilent Technologies at one of the following telephone  
numbers for warranty, service, or technical support information.  
In the United States: (800) 829-4444  
NOTE  
In Europe: 31 20 547 2111  
In Japan: 0120-421-345  
Or use our Web link for information on contacting Agilent in your country or  
Or contact your Agilent Technologies Representative.  
The web contains the most up to date version of the manual. Go to  
http://www.agilent.com/find/N5700 to get the latest version of the manual.  
4
Series N5700 User’s Guide  
Contents  
1 Quick Reference 7  
The Agilent N5700 DC Power Supplies – At a Glance 8  
The Front Panel - At a Glance 10  
The Rear Panel – At a Glance 12  
SCPI Programming Commands – At a Glance. 15  
2 Installation  
17  
General Information 18  
Inspecting the Unit 19  
Installing the Unit 19  
Connecting the Line Cord 21  
Connecting the Load23  
Output Voltage Sensing 26  
Load Considerations 28  
Parallel Connections30  
Series Connections 32  
J1 Connector Connections 34  
3 Operating the Power Supply Locally  
Turn-On Check-Out 36  
Normal Operation 38  
35  
Protection Functions39  
Output On/Off Control 42  
Analog Programming of Output Voltage and Current 44  
4 Operating the Power supply Remotely  
49  
Configuring the Interface50  
SCPI Commands – an Introduction 59  
5 Language Reference  
65  
Calibration Commands 66  
Measure Commands67  
Output Commands68  
Source Commands69  
Status Commands 71  
System Commands 77  
Trigger Commands 79  
Series N5700 User’s Guide  
5
6 Programming Examples 81  
Output Programming Example 82  
Trigger Programming Example 83  
A Specifications 85  
Performance Specifications 86  
Supplemental Characteristics 87  
Outline Diagram 89  
B Verification and Calibration  
91  
Verification 92  
Calibration 111  
C Service 113  
Types of Service Available 114  
Repackaging for Shipment 114  
Operating Checklist 114  
Error Messages 116  
D Compatibility 121  
Differences – In General 122  
Compatibility Command Summary 123  
Index 125  
6
Series N5700 User’s Guide  
1
Quick Reference  
The Agilent N5700 DC Power Supplies – At a Glance 8  
The Front Panel - At a Glance 10  
The Rear Panel – At a Glance 12  
SCPI Programming Commands – At a Glance. 15  
This chapter concisely describes the Agilent Technologies Series  
N5700 Power Supplies.  
This chapter is not meant to describe every operating feature in  
detail. It is simply a quick reference guide to quickly become familiar  
with the essential components of the power supply. It can also be  
used as a memory jogger for experienced users to quickly find a  
front/rear panel function or programming command.  
Series N5700 User’s Guide  
7
The Agilent N5700 DC Power Supplies – At a Glance  
The Agilent N5700 DC Power Supplies – At a Glance  
The Agilent Technologies Series N5700 System DC Power Supplies  
are general-purpose switching power supplies with a wide variety of  
output voltage and current ratings.  
These power supplies are power-factor corrected and operate from a  
worldwide AC voltage range. Output voltage and current are  
continuously displayed and LED indicators show the complete  
operating status of the power supply.  
The front panel controls allow the user to set the output parameters,  
over-voltage, under-voltage, and over-current protection levels, and  
preview the settings.  
The rear panel includes the necessary connectors to control and  
monitor the power supply operation by analog signals or by the built-  
in remote communication interfaces.  
Output Features  
Constant voltage/constant current with automatic crossover.  
High-resolution voltage and current front panel controls.  
Accurate voltage and current readback.  
Independent edge-triggered external shut-off, and level-  
triggered external enable/disable.  
Parallel master/slave operation with active current sharing.  
Remote sensing to compensate for voltage drop in load leads.  
Analog output programming and monitoring.  
System Features  
Built-in GBIB/LAN/USB interface.  
A built-in Web server that lets you control the instrument  
directly from an internet browser on your computer.  
Zero-gap stacking - no ventilation holes at the top and bottom  
surface of the power supply.  
Universal input voltage with active power factor correction.  
Fan speed control for low noise and extended fan life.  
8
Series N5700 User’s Guide  
The Agilent N5700 DC Power Supplies – At a Glance  
Programmable Functions  
Output voltage and current setting.  
Output voltage and current measurement.  
Output voltage and current trigger setting.  
Output On/Off control.  
Over-current protection setting.  
Over-voltage protection setting and readback.  
Under-voltage limit setting and readback.  
Start-up mode (either last setting or reset mode)  
Status register setting and readback.  
Bus trigger  
Calibration  
Model Ratings  
Model  
Voltage  
Range  
Current  
Range  
Model  
Voltage  
Range  
Current  
Range  
N5741A  
N5742A  
N5743A  
N5744A  
N5745A  
N5746A  
N5747A  
N5748A  
N5749A  
N5750A  
N5751A  
N5752A  
0 – 6V  
0 – 100A  
0 – 90A  
0 – 60A  
0 – 38A  
0 – 25A  
0 – 19A  
0 – 12.5A  
0 – 9.5A  
0 – 7.5A  
0 – 5A  
N5761A  
N5762A  
N5763A  
N5764A  
N5765A  
N5766A  
N5767A  
N5768A  
N5769A  
N5770A  
N5771A  
N5772A  
0 – 6V  
0 – 180A  
0 – 165A  
0 – 120A  
0 – 76A  
0 – 50A  
0 – 38A  
0 – 25A  
0 – 19A  
0 – 15A  
0 – 10A  
0 – 5A  
0 – 8V  
0 – 8V  
0 – 12.5V  
0 – 20V  
0 – 30V  
0 – 40V  
0 – 60V  
0 – 80V  
0 – 100V  
0 – 150V  
0 – 300V  
0 – 600V  
0 – 12.5V  
0 – 20V  
0 – 30V  
0 – 40V  
0 – 60V  
0 – 80V  
0 – 100V  
0 – 150V  
0 – 300V  
0 – 600V  
0 – 2.5A  
0 – 1.3A  
0 – 2.5A  
Series N5700 User’s Guide  
9
The Front Panel - At a Glance  
The Front Panel - At a Glance  
3
6
2
4
5
1
DC VOLTS  
DC AMPS  
VOLTAGE  
CV  
CURRENT  
CC  
OVP  
UVL  
LIMIT/  
OCP  
REM OUT ON  
PROT  
FINE  
POWER  
19  
16  
15  
14  
11  
7
13  
9
18  
8
17  
10  
12  
1 – VOLTAGE knob  
Voltage function: Adjusts the output voltage, the over-voltage protection level,  
and the under-voltage limit. If over-voltage protection or under-voltage limits have  
been set, you cannot program the output voltage outside those limits.  
GPIB address: Selects the GPIB address when REM is pressed and held.  
2 – CV indicator  
When lit, indicates that the unit is operating in constant voltage mode – with the  
output voltage being held constant.  
3 – DC VOLTS display  
LED display that normally displays the voltage measured at the sense terminals.  
When LIMIT is pressed, the display indicates the programmed voltage setting.  
When OVP/UVL is pressed, the display indicates either the OVP or UVL setting.  
When REM is pressed and held, the display indicates the GPIB address.  
4 – DC AMPS display  
5 – CC indicator  
LED display that normally displays the current measured at the output terminals.  
When LIMIT is pressed, the display indicates the programmed current setting.  
When lit, indicates that the unit is operating in constant current mode – with the  
output current being held constant.  
6 – CURRENT knob  
7 – OUT ON button  
Adjusts the output current.  
Output function: Press OUT ON to turn the output on or off. Press OUT ON to  
reset and turn the output on after an OVP or OCP event has occurred.  
Start-up function: Selects between Safe-Start and Auto-Restart modes. Press and  
hold the OUT ON button to toggle between Safe-Start and Auto-Restart. The  
display cycles between SAF and AU7. Releasing the OUT ON button while one of  
the modes is displayed selects that mode.  
8 – OUT ON indicator  
When lit, indicates that the output is enabled or on.  
10  
Series N5700 User’s Guide  
The Front Panel - At a Glance  
9 – REM button  
Mode function: Press REM to put the unit into local mode. (This button can be  
disabled with a Local Lockout command).  
Address function: Selects the GPIB address. Press and hold the REM button for  
three seconds to set the address with the Voltage knob.  
10 – REM indicator  
11 – OCP button  
When lit, indicates that the unit is in Remote mode.  
Enable function: Press OCP to turn over-current protection on. Press OCP again to  
turn over-current protection off.  
Reset OCP: When an over-current protection event occurs, press the OUT ON  
button to enable the output and re-arm over-current protection.  
12 – OCP indicator  
When lit, indicates that over-current protection is enabled or on.  
13 – OVP/UVL button  
OVP function: Press OVP/UVL once to set the over-voltage protection level with  
the Voltage knob (the display shows OUP). You cannot set the over-voltage  
protection lower than about 5% above the present output voltage setting.  
UVL function: Press OVP/UVL twice to set the under-voltage programming limit  
with the Voltage knob (the display shows UUL). You cannot set the under-voltage  
protection higher than about 5% below the present output voltage setting.  
14 – LIMIT button  
Limit function: Press LIMIT to display the output voltage and current limit. For five  
seconds the display shows the settings and then it returns to show the actual  
output voltage and current.  
Lock function: Press and hold the LIMIT button to toggle between Locked front  
panel and Unlocked front panel. The display will cycle between LFP and UFP.  
Releasing the LIMIT button while one of the modes is displayed selects that  
mode.  
15 – LIMIT indicator  
16 – FINE button  
When lit, indicates that the LIMIT button is pressed.  
Selects Fine or Coarse adjustment control. In Fine mode, the Voltage and Current  
knobs operate with high resolution; in Coarse mode, with lower resolution  
(approximately six turns).  
17 – FINE indicator  
18 – PROT indicator  
When lit, indicates that the unit is in Fine adjustment mode.  
When blinking, indicates that a fault has occurred.  
OVP, OCP, OTP, Enable fail, and AC fail detection will cause the PROT indicator to  
blink. The PROT indicator may blink and the display indicate AC for a few seconds  
after the unit is turned off because of residual energy inside the unit.  
19 – POWER switch  
Turns the power supply on or off.  
Series N5700 User’s Guide  
11  
The Rear Panel – At a Glance  
The Rear Panel – At a Glance  
8
9
10/100 Ethernet  
!
!
LINK  
TX  
+V  
-V  
GPIB  
J1  
SW1  
ON  
J2  
OFF  
ANALOG PROGRAMMING  
1 2 3 4 5 6 7 8 9  
+S+LS NC -LC-S  
NOT ACTIVE  
AC INPUT  
80V - 600V  
750W  
2
1
7
6
3
5
4
6V - 60V  
1500W  
1 – AC input connector  
2 – DC output connector  
Wire clamp connector for 1500W output models.  
IEC connector for 750W output models.  
Wire clamp connector for 80V to 600V models.  
Bus bars for 6V to 60V models.  
3 – USB connector  
4 – LAN connector  
Connector for connecting to a USB interface. See chapter 4 for setup.  
Connector for connecting to a LAN interface. LINK LED indicates link integrity.  
TX LED indicates LAN activity. See chapter 4 for LAN setup.  
5 – Analog Programming  
connector  
Connector for the analog interface. Includes output voltage and current limit  
programming and monitoring signals, Shut-Off control (electrical signal),  
Enable/Disable control (dry-contact), power supply ok (Power Supply OK) signal  
and operation mode (CV/CC) signal. (See next page for details)  
6 – SW1 setup switch  
Nine-position switch for selecting remote programming and monitoring modes  
for Output Voltage, Current Limit and other control functions. (See next page for  
details)  
7 – Remote Sense connector Connector for making remote sensing connections for regulating the load  
voltage and compensating for wiring voltage drop. (See next page for details)  
8 – GPIB connector  
9 – Ground screw  
Connector for connecting to a GPIB interface. See chapter 4 for setup.  
M4x8 screws for making chassis ground connections  
SHOCK HAZARD The power cord provides a chassis ground through a third  
conductor. Be certain that your power outlet is of the three-conductor type  
with the correct pin connected to earth ground  
WARNING  
12  
Series N5700 User’s Guide  
The Rear Panel – At a Glance  
J2 Sense Connector  
1 – Remote sense (+)  
2 – Local sense (+)  
3 – Not used  
4 – Local sense (–)  
5 – Remote sense (–)  
The factory-shipped configuration is shown in the figure.  
SW1 Setup Switch  
9
8
7
6
5
4
3
2
1
The factory-shipped setting is Down for all switches.  
1 – Output voltage, voltage  
programming  
Down: The output voltage is programmed by the front panel.  
Up:  
The output voltage is programmed by the external voltage signal.  
2 – Output current, voltage  
programming  
Down: The output current is programmed by the front panel.  
Up: The output current is programmed by the external voltage signal.  
3 – Programming range  
(voltage/resistance)  
Down: The remote programming range is: 0 – 5V / 0 – 5K.  
Up: The remote programming range is: 0 – 10V / 0 – 10K.  
4 – Voltage and Current  
monitoring range  
Down: The remote monitoring range is: 0 – 5V.  
Up: The remote programming range is: 0 – 10V.  
5 – Shut-Off Logic Select  
Down: OUT OFF = Low (0 – 0.6V) or short; OUT ON = High (2V – 15V) or open.  
Up: OUT OFF = High (2V – 15V) or open; OUT ON = Low (0 – 0.6V) or short.  
6 – Not Used  
7 – Output voltage, resistive Down: The output voltage is programmed by the front panel.  
programming Up: The output voltage is programmed by the external resistor.  
8 – Output current, resistive Down: The output current is programmed by the front panel.  
programming  
Up:  
The output current is programmed by the external resistor.  
9 – Enable/Disable control  
Down: The J1 Enable+/Enable– pins are not active.  
Up: The J1 Enable+/Enable– pins are active.  
Series N5700 User’s Guide  
13  
The Rear Panel – At a Glance  
J1 Analog Programming Connector  
Current Program  
Voltage Program  
Local / Analog  
Voltage Monitor  
Common (-S)  
CV / CC  
Chassis Common  
Chassis Common  
Enable +  
13  
25  
12  
24  
11  
23  
10  
22  
9
8
7
6
5
4
3
2
1
21  
20  
19  
18  
17  
16  
15  
14  
Parallel  
Enable --  
Current Monitor  
Current Prog. Return  
Voltage Prog. Return  
Local / Analog State  
Shut Off  
Power Supply OK  
The factory-shipped default configuration is Local operation, which  
does not require connection to J1.  
Pin 1:  
Enable +  
Connect Pin 1 to Pin 14 to enable the output. Disconnect to disable the output.  
Signal return for Pin 15 and Pin 16. Connected to chassis.  
No connection  
Pin 2, 3:  
Pin 4–7:  
Pin 8:  
Chassis Common  
Not Used  
Local/Analog  
Voltage Program  
Current Program  
Voltage Monitor  
Common  
Input for selecting between front panel or analog programming of the output.  
Input for voltage or resistance programming of the output voltage.  
Input for voltage or resistance programming of the output current.  
Output for monitoring the output voltage.  
Pin 9:  
Pin 10:  
Pin 11:  
Pin 12:  
Pin 13:  
Pin 14:  
Pin 15:  
Pin 16:  
Signal return for Pin 8, Pin11, Pin 13, and Pin 24. Connected internally to –S.  
Output for constant voltage/constant current mode indication.  
Connect Pin 14 to Pin 1 to enable the output. Disconnect to disable the output.  
Input for Shut-Off control of the output. Referenced to Chassis Common.  
Output to indicate the power supply status. Referenced to Chassis Common.  
No connection  
CV/CC  
Enable –  
Shut Off  
Power Supply OK  
Pin 17–20: Not Used  
Pin 21:  
Pin 22:  
Pin 23:  
Pin 24:  
Pin 25:  
Local/Analog State  
Output for indication of local or analog programming mode.  
Signal return for Pin 9. Connected internally to –S.  
Voltage Prog. Return  
Current Prog. Return  
Current Monitor  
Parallel  
Signal return for Pin 10. Connected internally to –S.  
Output for monitoring the output current.  
Output for current balancing in parallel operation.  
14  
Series N5700 User’s Guide  
SCPI Programming Commands – At a Glance.  
SCPI Programming Commands – At a Glance.  
Some [optional] commands have been included for clarity. Chapter 6 contains a  
complete description of all programming commands.  
NOTE  
Subsystem Commands.  
SCPI Command  
Description  
ABORt  
Aborts the triggered action  
CALibrate  
:CURRent[:LEVel]  
:DATA <NRf>  
:DATE “<date>”  
:LEVel P1 | P2  
:PASSword <NRf>  
:STATE <Bool> [,<NRf>]  
:VOLTage[:LEVel]  
Calibrates the output current programming  
Enters the calibration value  
Sets the calibration date  
Advances to the next calibration step  
Sets the numeric calibration password  
Enables/disables calibration mode  
Calibrates the output voltage programming  
INITiate  
[:IMMediate][:TRANsient]  
Initiates the trigger system  
:CONTinuous[:TRANsient]  
Enables/disables continuous triggers  
MEASure  
[:SCALar]  
:CURRent[:DC]?  
:VOLTage[:DC]?  
Returns the measured output current  
Returns the measured output voltage  
OUTPut  
[:STATe] <Bool>  
:PON  
:STATe RST | AUTO  
:PROTection  
:CLEar  
Enables/disables the specified output  
Programs the Power-On State  
Resets latched protection  
[SOURce:]  
CURRent  
[:LEVel]  
[:IMMediate][:AMPLitude] <NRf+>  
:TRIGgered[:AMPLitude] <NRf+>  
:PROTection  
:STATe <Bool>  
VOLTage  
[:LEVel]  
Sets the output current  
Sets the triggered output current  
Enables/disables over-current protection  
[:IMMediate][:AMPLitude] <NRf+>  
:TRIGgered[:AMPLitude] <NRf+>  
Sets the output voltage  
Sets the triggered output voltage  
:LIMit  
:LOW <NRf+>  
:PROTection  
Sets the low-voltage limit  
[:LEVel] <NRf+>  
Sets the over-voltage protection level  
Series N5700 User’s Guide  
15  
SCPI Programming Commands – At a Glance.  
SCPI Command  
Description  
STATus  
:OPERation  
[:EVENt]?  
:CONDition?  
:ENABle <NRf>  
:NTRansition<NRf>  
:PTRansition<NRf>  
:PRESet  
Returns the value of the operation event register  
Returns the value of the operation condition register  
Enables specific bits in the Event register  
Sets the Negative transition filter  
Sets the Positive transition filter  
Presets all enable and transition registers to power-on  
:QUEStionable  
[:EVENt]?  
Returns the value of the questionable event register  
Returns the value of the questionable condition register  
Enables specific bits in the Event register  
Sets the Negative transition filter  
:CONDition?  
:ENABle <NRf>  
:NTRansition<NRf>  
:PTRansition<NRf>  
Sets the Positive transition filter  
SYSTem  
:COMMunicate  
:RLSTate LOCal | REMote | RWLock  
:ERRor?  
Specifies the Remote/Local state of the instrument  
Returns the error number and error string  
Returns the SCPI version number  
:VERSion?  
TRIGger  
:SOURce BUS  
[:TRANsient][:IMMediate]  
Sets the measurement trigger source  
Generates a transient trigger  
Common Commands  
Command  
Description  
*CLS  
Clear status  
*ESE <NRf>  
*ESE?  
*ESR?  
*IDN?  
*OPC  
*OPC?  
*OPT?  
Standard event status enable  
Return standard event status enable  
Return event status register  
Return instrument identification  
Enable "operation complete" bit in ESR  
Return a "1" when operation complete  
Return option number  
*RCL <NRf>  
*RST  
*SAV <NRf>  
*SRE <NRf>  
*SRE?  
Recalls a saved instrument state  
Reset  
Saves an instrument state  
Set service request enable register  
Return service request enable register  
Return status byte  
*STB?  
*TRG  
*TST  
*WAI  
Trigger  
Performs self-test, then returns result  
Holds off bus until all device commands done  
16  
Series N5700 User’s Guide  
2
Installation  
General Information 18  
Inspecting the Unit 19  
Installing the Unit 19  
Connecting the Line Cord 21  
Connecting the Load23  
Output Voltage Sensing 26  
Load Considerations 28  
Parallel Connections30  
Series Connections 32  
J1 Connector Connections 34  
This chapter describes how to install your power supply. It discusses  
installation, rack mounting, and line cord connections.  
This chapter also discusses how to connect your load to the output  
terminals. It discusses what you need to know about wire sizes and  
how to compensate for voltage drops in the load leads. It also  
discusses various loads configurations and how to connect units in  
series and parallel.  
Before getting started, check the list under “Items Supplied” and  
verify that you have received these items with your instrument. If  
anything is missing, please contact your nearest Agilent Sales and  
Service Office.  
Series N5700 User’s Guide  
17  
General Information  
General Information  
Models  
750 W Models  
1500 W Models  
N5741A – N5749A  
N5750A – N5752A  
N5761A – N5769A  
N5770A – N5772A  
Options  
Option  
Description  
Description  
Accessories  
Item  
N5740A  
Rack-mount Slide Kit for installing in system II style cabinets  
Items Supplied  
Item  
Description  
Power Cord  
A power cord appropriate for your location.  
750W units are supplied with terminated power cords  
1500W units are supplied with unterminated power cords  
Sense Connector  
Analog connector  
A 5-pin connector for local/remote sense connections  
A DB25 subminiature connector plug for analog control  
connections  
Shield assembly  
A safety shield for the output terminal connections  
Strain relief assembly  
A strain relief assembly for unterminated power cords  
(only provided for 1500W units)  
Documentation Set  
Contains User’s Guide and CD-ROM  
CD-ROM contains Agilent I/O library, Setup utility,  
GPIB/LAN/USB Interfaces Connectivity Guide, User’s Guide  
18  
Series N5700 User’s Guide  
Inspecting the Unit  
Inspecting the Unit  
Installing the Unit  
When you receive your power supply, inspect it for any obvious  
damage that may have occurred during shipment. If there is damage,  
notify the shipping carrier and nearest Agilent Sales and Service  
Office immediately. Refer to Appendix C for more information.  
Until you have checked out the power supply, save the shipping  
carton and packing materials in case the unit has to be returned.  
Safety Considerations  
This power supply is a Safety Class 1 instrument, which means it has  
a protective earth terminal. That terminal must be connected to  
earth ground through power source equipped with a ground  
receptacle. Refer to the Safety Summary page at the beginning of this  
guide for general safety information. Before installation or  
operation, check the power supply and review this guide for safety  
warnings and instructions. Safety warnings for specific procedures  
are located at appropriate places throughout this Guide.  
Environment  
Do not operate the instrument in the presence of flammable gasses or fumes  
WARNING  
The environmental conditions, dimensions of the instrument, as well  
as an outline diagram are given in Appendix A. Basically, the  
instrument should only be operated indoors in a controlled  
environment. Do not operate the power supply in an area where the  
ambient temperature exceeds 40° C.  
Agilent N5700 power supplies generate magnetic fields, which may affect the  
operation of other instruments. If your equipment is susceptible to magnetic  
fields, do not position it adjacent to the power supply.  
NOTE  
Airflow  
Fans cool the power supply by drawing air through the front and  
exhausting it out the back. The instrument must be installed in a  
location that allows sufficient space of at least 10 cm (4 in) at the  
front and back of the unit for adequate air circulation.  
Series N5700 User’s Guide  
19  
Installing the Unit  
Rack Installation  
Ensure that the screws used to attach the rack slide kit do not penetrate more  
than 6 mm into the sides of the unit.  
CAUTION  
Do not block the air intake at the front of the unit or the exhaust at the rear of  
the unit.  
The Agilent N5700 power supplies can be mounted in a standard 19-  
inch rack panel or cabinet. To install the power supply in a rack:  
1. Use the front panel rack-mount brackets to install the power  
supply in the rack.  
2. Use a support bracket to provide adequate support for the rear of  
the power supply.  
3. If using rack mount slides, use Agilent N5740A Rack-mount Slide  
Kit to install the unit in a standard 19-inch equipment rack. Refer to  
the following figure for assembly instructions. Use three #10-32 x  
0.38 in (max.) screws at each side. To prevent internal damage, use  
the specified screw length only.  
Cleaning  
SHOCK HAZARD To prevent electric shock, unplug the unit before cleaning.  
WARNING  
Use a dry cloth or one slightly dampened with water to clean the  
external case parts. Do not attempt to clean internally.  
20  
Series N5700 User’s Guide  
Connecting the Line Cord  
Connecting the Line Cord  
SHOCK HAZARD The power cord provides a chassis ground through a third  
conductor. Be certain that your power outlet is of the three-conductor type  
with the correct pin connected to earth ground.  
WARNING  
FIRE HAZARD Use only the power cord that was supplied with your  
instrument. Using other types of power cords may cause overheating of the  
power cord, resulting in fire.  
The detachable power cord may be used as an emergency disconnecting  
device. Removing the power cord will disconnect ac input power to the unit.  
NOTE  
The AC input on the back of your unit is a universal AC input. It  
accepts line voltages in the range of 85 VAC to 265 VAC. The  
frequency range is 47 Hz to 63 Hz.  
The input current requirement of 750W units is 10.5A @ 100 VAC  
nominal and 5A @ 200 VAC nominal. The current requirement of  
1500W units is 21A @ 100 VAC nominal and 11A @ 200 VAC nominal.  
Input Connections for 750W units  
Connect the power cord to the IEC 320 connector on the rear of the  
unit. The IEC connector provides the safety ground connection when  
the AC cord is plugged into a grounded AC receptacle.  
If the wrong power cord was shipped with your unit, contact your  
nearest Agilent Sales and Service Office.  
Input Connections for 1500W units  
Connection of this power supply to an AC power source should be made by a  
qualified electrician or other qualified personnel.  
CAUTION  
The AC input connector is a 3-terminal wire clamp located on the  
rear panel. Use suitable wires and tightening torque as follows:  
Wire diameter: 12 AWG or 10 AWG  
Tightening torque: 6.5 - 7.0 in-lb  
Connect the cable to the AC input connector as follows:  
Strip the outside insulation of the AC cable approximately 10  
cm (4 in). Trim the wires so that the ground wire is 10 mm  
(0.4 in) longer than the other wires. Strip 14 mm (0.55 in) at  
the end of each of the wires.  
Series N5700 User’s Guide  
21  
Connecting the Line Cord  
Unscrew the base of the strain relief from the helix-shaped  
body. Insert the base through the outside opening in the AC  
input cover and screw the locknut securely (11-14 in-lb) into  
the base, from the inside.  
Slide the helix-shaped body onto the AC cable. Insert the  
stripped wires through the strain relief base until the outer  
cable jacket is flush with the edge of the base. Tighten the  
body (16-18 in-lb) to the base while holding the cable in  
place. Now the cable is securely fastened inside the strain  
relief. Refer to the following figure.  
Screw-on  
Locknut  
Route the AC wires to the input connector terminals as  
required. To connect the wires, loosen the terminal screw,  
insert the stripped wire into the terminal, and tighten the  
screw securely to between 4.4–5.3 in-lb.  
Route the wires inside the cover to prevent pinching. Fasten  
the cover to the unit using the M3 x 8 flat head screws  
provided. Refer to the following figure for details.  
N
L
M3x8  
Flat Head Screws  
(2 places)  
Cover  
Assembled  
Strain Relief  
22  
Series N5700 User’s Guide  
Connecting the Load  
Connecting the Load  
SHOCK HAZARD Turn off AC power before making rear panel connections.  
All wires and straps must be properly connected with screws securely  
tightened.  
WARNING  
As further explained in this section, the following factors should be  
considered when selecting wiring to connect the load to the power  
supply:  
Current carrying capacity of the wire  
Insulation rating of the wire should be at least equivalent to  
the maximum output voltage of the power supply  
Maximum wire length and voltage drop  
Noise and impedance effects of the load wiring  
Wire Size  
FIRE HAZARD To satisfy safety requirements, select a wire size heavy  
enough not to overheat while carrying the power supply load current at the  
rated load, or the current that would flow in the event the load wires were  
shorted, whichever is greater.  
WARNING  
Along with conductor temperature, you must also consider voltage  
drop when selecting wire sizes. The following chart lists the  
resistance for various wire sizes and also the maximum lengths to  
limit the voltage drop to 1.0 volt for various currents.  
Although the power supply will compensate for up to 5V in each load  
wire, it is recommended to minimize the voltage drop to less than 1V  
to prevent excessive output power consumption from the power  
supply and poor dynamic response to load changes.  
Wire size Resistance  
Maximum length in feet to limit voltage to 1 V  
AWG  
/1000 foot  
for 5 A  
for 10 A for 20A  
for 50A for 150A  
14  
12  
10  
8
6
4
2.526  
1.589  
80  
40  
60  
20  
30  
50  
80  
125  
200  
300  
500  
8
2
3.4  
6
10  
16  
26  
40  
68  
120  
200  
320  
500  
800  
1200  
2000  
12  
20  
32  
50  
80  
125  
200  
0.9994  
0.6285  
0.3953  
0.2486  
0.1564  
0.0983  
100  
160  
250  
400  
600  
1000  
2
0
Series N5700 User’s Guide  
23  
Connecting the Load  
Cross  
section  
(mm2)  
Resistance  
/kilometer  
Maximum length in meters to limit voltage to 1 V  
for 5 A for 10 A  
for 20A  
for 50A for 150A  
2.5  
4
6
10  
16  
25  
35  
8.21  
5.09  
3.39  
1.95  
1.24  
0.795  
0.565  
24.0  
39.2  
59.0  
102  
160  
250  
354  
12.0  
18.6  
29.4  
51.2  
80.0  
125  
6.0  
9.8  
2.4  
4.0  
5.8  
10.2  
16.0  
25.2  
35.4  
0.8  
1.4  
2.0  
3.4  
5.4  
8.4  
11.8  
14.8  
25.6  
40.0  
62.0  
88.0  
177  
Load Connections for 6V to 60V Models  
SHOCK HAZARD Hazardous voltages may exist at the outputs and the load  
connections when using a power supply with a rated output greater than 40V.  
To protect personnel against accidental contact with hazardous voltages,  
ensure that the load and its connections have no accessible live parts. Ensure  
that the load wiring insulation rating is greater than or equal to the maximum  
output voltage of the power supply.  
WARNING  
Ensure that the load wiring mounting hardware does not short the output  
terminals. Heavy connecting cables must have some form of strain relief to  
prevent loosening the connections or bending the bus-bars.  
CAUTION  
All load wires should be properly terminated with wire terminals  
securely attached. Do not use unterminated wires for load  
connections at the power supply. The following figures illustrate how  
to connect the load wires to the power supply bus-bars as well as  
how to mount the bus-bar shield to the chassis.  
Wire terminal lug (2 places)  
M8x15 screw (2 places)  
Flat washer  
(2 places)  
Flat washer (2 places)  
Spring washer (2 places)  
Hex Nut (2 places)  
Screws tightening torque: 104-118 in-lb.  
24  
Series N5700 User’s Guide  
Connecting the Load  
Install the shield after you have finished connecting the load wires.  
Shield  
Load Connections for 80V to 600V Models  
SHOCK HAZARD Hazardous voltages may exist at the outputs and the load  
connections when using a power supply with a rated output greater than 40V.  
To protect personnel against accidental contact with hazardous voltages,  
ensure that the load and its connections have no accessible live parts. Ensure  
that the load wiring insulation rating is greater than or equal to the maximum  
output voltage of the power supply.  
WARNING  
The 80V to 600V models have a four-terminal wire clamp output  
connector. The two left terminals are the positive outputs and the  
two right terminals are the negative outputs. The connector  
specifications are as follows:  
Wire Size:  
Stripping Length: 10 mm (0.39 in.)  
Torque: 6.5 - 7 in-lb.  
AWG 18 to AWG 10  
The following instructions describe how to connect the load wires to  
the power supply:  
Strip wires back approximately 10 mm (0.39 in).  
Loosen the connector terminal screws and insert the stripped  
wires into the terminal. Tighten the terminal screw securely.  
-V  
+V  
Positive Output (+)  
Negative (-)  
Output/Return  
Load wires  
Series N5700 User’s Guide  
25